Quantum Instruments Cascade Laser Starter Kit Manual do Utilizador Página 36

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 35
Chapter 7
Troubleshooting
7.1 Chapter overview
This chapter sets out troubleshooting instructions
for ensuring safe and trouble-free operation of the
Quantum Cascade Laser Starter Kit system.
7.2 Principle
The personnel must have read and understood this
documentation before carrying out any activity what-
soever with the Quantum Cascade Laser Starter
Kit. In ca se of unclear information, please contact
the distributor.
The power supply reaches max cur-
rent for a very low voltage
- The low impedance line or the QCL is
shorted to ground.
Check the cabling.
Replace the laser if necessary.
- The DC power supply is NOT floating.
Change the cabling in order to leave
the laser floating.
- The laser tightening device is too tight-
ened and the QCL lase r is shorted to
ground.
Loose the laser fixture device.
35
Vista de página 35
1 2 ... 31 32 33 34 35 36 37 38 39 40 41 ... 59 60

Comentários a estes Manuais

Sem comentários